Fluke Calibration CIT MET/CAL Procedure ============================================================================= INSTRUMENT: Validyne Model P55D 8.0 PSI (Pos-Neg) with PPC2+ DATE: 2019-10-21 17:10:32 AUTHOR: Customer REVISION: 1.2 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 11 NUMBER OF LINES: 116 CONFIGURATION: HP 3458A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # 1.001 RSLT = T.O. 33K6-4-1425-1 Technical Manual # # **************************************************************************** 1.002 HEAD ENVIRONMENTAL CONDITIONS # **************************************************************************** 1.003 DISP Ensure the UUT has been allowed to stabilize in an 1.003 DISP environment with an ambient temperature of 18 to 28 1.003 DISP Degrees C and a relative humidity of < 90%. # # **************************************************************************** 1.004 HEAD INITIAL CONDITIONS & UUT HOOKUP # **************************************************************************** # 1.005 ASK+ U Z 1.006 ASK- R N P A W 1.007 VSET NMEAS = 1 # Auto assign 1.008 HEAD TEST SET UP PROCEDURE 1.009 DISP 1. Connect DHI test port to transducer inlet (+). 1.009 DISP 2. Ensure that pressure and vacuum supplies are on. 1.009 DISP 3. Ensure that the transducer under test is +/- 8.0 PSI. 1.009 DISP (CODE 38) 1.009 DISP 4. Attach connector to transducer. 1.009 DISP 5. Turn on DC power supply and adjust to +15 VDC. 1.010 RESET [@PPC2]VENT=1[D1000] 1.011 MATH @DwellTime = 10 1.012 HEAD ZERO TEST 1.013 TSET TDESC = Nominal @ 0.00 PSI 1.014 IEEE2 [@PPC2]MSGFMT=0 1.015 IEEE2 [@PPC2]RANGE=2:LO 1.016 IEEE2 [@PPC2]UNIT=PSIG 1.017 IEEE2 [@PPC2]MODE=1 1.018 IEEE2 [@PPC2]HS=.008 1.019 IEEE2 [@PPC2]PS=0.00 1.020 CALL Sub wait for ready state with PPC2+ 1.021 DISP Using 'Z' pot, adjust output to 0 Volts. 1.022 3458 10 0.000V 0.025U 2W 2.001 HEAD 1.25 VOLT TEST 2.002 TSET TDESC = Nominal @ 2.00 PSI 2.003 IEEE2 [@PPC2]PS=2.0 2.004 CALL Sub wait for ready state with PPC2+ 2.005 3458 10 1.250V 0.025U 2W 3.001 HEAD 2.5 VOLT TEST 3.002 TSET TDESC = Nominal @ 4.00 PSI 3.003 IEEE2 [@PPC2]PS=4.0 3.004 CALL Sub wait for ready state with PPC2+ 3.005 3458 10 2.500V 0.025U 2W 4.001 HEAD 3.75 VOLT TEST 4.002 TSET TDESC = Nominal @ 6.00 PSI 4.003 IEEE2 [@PPC2]PS=6.0 4.004 CALL Sub wait for ready state with PPC2+ 4.005 3458 10 3.750V 0.025U 2W 5.001 HEAD 5 VOLT TEST 5.002 TSET TDESC = Nominal @ 8.00 PSI 5.003 IEEE2 [@PPC2]PS=8.0 [D30000] 5.004 CALL Sub wait for ready state with PPC2+ 5.005 3458 10 5.000V 0.025U 2W 6.001 HEAD ZERO2 TEST 6.002 TSET TDESC = Nominal @ 8.00 PSI 6.003 IEEE2 [@PPC2]PS=0 6.004 CALL Sub wait for ready state with PPC2+ 6.005 3458 10 0.000V 0.025U 2W # SET RANGE TO HI RANGE MODULE SLOT 5 7.001 IEEE2 [@PPC2]RANGE=1:HI 7.002 HEAD -1.25 VOLT TEST 7.003 TSET TDESC = Nominal @ -2.00 PSI 7.004 IEEE2 [@PPC2]PS=-2.0 7.005 CALL Sub wait for ready state with PPC2+ 7.006 3458 10 -1.250V 0.025U 2W 8.001 HEAD -2.5 VOLT TEST 8.002 TSET TDESC = Nominal @ -4.00 PSI 8.003 IEEE2 [@PPC2]PS=-4.0 8.004 CALL Sub wait for ready state with PPC2+ 8.005 3458 10 -2.500V 0.025U 2W 9.001 HEAD -3.75 VOLT TEST 9.002 TSET TDESC = Nominal @ -6.00 PSI 9.003 IEEE2 [@PPC2]PS=-6.0 9.004 CALL Sub wait for ready state with PPC2+ 9.005 3458 10 -3.750V 0.025U 2W 10.001 HEAD -5 VOLT TEST 10.002 TSET TDESC = Nominal @ -8.00 PSI 10.003 IEEE2 [@PPC2]PS=-8.0 10.004 CALL Sub wait for ready state with PPC2+ 10.005 3458 10 -5.000V 0.025U 2W 11.001 IEEE2 [@PPC2]VENT=1 11.002 3458 * S 11.003 END