Begin Header instrument = Fluke 5128A interval = 365 days confidence = 2.58 sigma End Header # $Revision: 35443 $ # Fluke 5128A RHapid-Cal Humidity Generator, Operators Manual, December 2016 # Humidity Not Used Tol (%) Floor (%RH) Resolution # ------------------------ ---------------- ------- ----------- ---------- Mode: Humidity UNC 7.00 80.00 NA NA 0 1.00 0.01 80.00 95.00 NA NA 0 1.25 0.01 Mode: Humidity Uniformity 7.00 95.00 NA NA 0 0.30 0.01 Mode: Humidity Stability 7.00 95.00 NA NA 0 0.10 0.01 # Temperature Not Used Tol (%) Floor (°C) Resolution # ------------------------ ---------------- ------- ----------- ---------- Mode: Temperature UNC 18.00 28.00 NA NA 0 0.20 0.01 Mode: Temperature Uniformity 18.00 28.00 NA NA 0 0.12 0.01 Mode: Temperature Stability 18.00 28.00 NA NA 0 0.05 0.01