Views:

See the attached document for seal kit and instructions to replace the seals. Seal replacement in the intensifier is quite difficult. It is recommended that this work only be done by Fluke Calibration or Fluke Calibration Authorized Service Providers. 

See this article for leak checking, P3800 Deadweight Tester Leak Check

4105694 Rev 003 P3800SK,SEAL KIT, P3860, P3840 & P3830
Includes these parts
PN Qty Description
3910866 1 72-39,SEAL .563 OD X .313 ID, POLYPAK
3868028 2 S51-0007,SEAL O-RING 1-1/4 X 1/8  90 DURO BUNA-N (NITRILE)
3923130 2 S51-0008,BACKUP RING,PTFE,SIZE BS218,SPLIT,1-1/4 X 1-1/2IN OD X .058 IN T,SPLIT
3923683 2 SDW4546,LENS RING, TOOL STEEL SEE GP SPECS
3921935 7 P109670,BONDED SEAL [PP-45-A-SC] [SELF CENTERING]
3924047 1 SR3603,SEAL WASHER (RUBBER) APROX .53IDX.81ODX.13THK
3924058 1 SR3604,BACK-UP RING PTFE BS.207
3924020 1 SR3585,O-RING REF.200-114- 4470
3862349 1 25-1056,FTG GASKET (BONDED SEAL)
3923940 1 SR3201,SEAL- BONDED 3/8 BSP EXT 400-023-4490-41
3887682 1 S51-0034,SEAL O-RING NITRILE SIZE BS615, SAME AS P8128
3922806 1 S421-202-76,WASHER 1/4IN BSP NYLON DRG.ISS.1
3922096 2 P109740,BONDED SEAL[PP45-B]VITON BSP

 

 