Views:

Please note this special statement from the MET/CAL 10.6 readme file:

mceclip1.png

You must re-configure your prompt files, either just in MET/CAL Runtime if you are using the default prompts or in both MET/CAL Runtime and MET/TEAM.

1 - If you only use the default prompt files, log in to Runtime, click  Configuration and go to Files and Folders.

Simply select the prompts you want to use from the ellipsis and click the Find button.  Then select the desired file from the list of active prompt files and click Apply.  That is all you have to do. 

 

2 - If you need to create a new prompt file from scratch or modify the stock versions, here are a couple of tips:

This process is done from MET/TEAM and you will need to have access to the configure menu.  

Configure>MET/CAL prompt files.  Then click search to see all of the available templates.

To add another field, simply right-click into the white space as shown below:

mceclip0.png

Then you can select fields to add from the drop-down list and customize them any way you need.  

Finally save and make active.  Then follow step 1 above to select your modified prompt files in your Runtime configuration. 

 

If you have further questions regarding prompt files or any of our Fluke Calibration Software, please contact softwaresupport@flukecal.com