Question:
- I can only measure the loop impedance between L and N, but no longer between L and PE. What can be the reason?
Answer:
You have probably accidentally activated the mode for testing in IT systems (also called IT network).
In this system the PE is not connected to the power supply, so a loop impedance measurement cannot be performed.
To switch back to normal test mode, please follow the steps below follows:
- Turn off the installation tester
- With the test instrument switched off, press and hold the F2 key on the installation tester
- While holding down the F2 key, switch on the installation tester
- You will now see “IT OFF” briefly in the display
Newer instruments have an additional option for IT mode, described on page 3 of the manual supplement.
You can recognize these newer instruments by the yellow silkscreen text next to the "up/down" and "light" keys.
- Press the backlight key for 1 second to see the IT mode status
- Use the Up/Down arrow key to toggle the IT mode On or Off
You should now be able to switch between L-N and L-PE again when measuring the loop impedance (with and without RCD)