Views:

Tech Notes

          

--------------------------------------------------------------------------------------------------------------------------------

Title:     5606 probe accuracies at -197°C, -38°C, and 100°C

Created:  8 Apr 2013

Last Revised:

--------------------------------------------------------------------------------------------------------------------------------

 

 

Applies To:  5606

 

Problem Description:  How can I find or determine the accuracies on the 5606?

 

Resolution/Work Around:   The accuracies can be determined by the following:

 

 

 

-197°C

-38°C

100°C

drift

0.0057

0.0252

0.0417

cal uncertainty

0.025

0.025

0.045

Sum (RSS)

0.000657

0.00126

0.003764

Accuracy °C=

0.026

0.035

0.06