Tech Notes
--------------------------------------------------------------------------------------------------------------------------------
Title: 5606 probe accuracies at -197°C, -38°C, and 100°C
Created: 8 Apr 2013
Last Revised:
--------------------------------------------------------------------------------------------------------------------------------
Applies To: 5606
Problem Description: How can I find or determine the accuracies on the 5606?
Resolution/Work Around: The accuracies can be determined by the following:
|
-197°C |
-38°C |
100°C |
drift |
0.0057 |
0.0252 |
0.0417 |
cal uncertainty |
0.025 |
0.025 |
0.045 |
Sum (RSS) |
0.000657 |
0.00126 |
0.003764 |
Accuracy °C= |
0.026 |
0.035 |
0.06 |