Parts that are needed:
Procedure to replace the Power Entry Module:
Before replacing the power entry module turn OFF the unit and unplug the power cord from the wall
1. Remove the right and left panels (10 screws per each side)
2. Remove the stir motor cover (2 screws)
3. Remove the access cover
4. Remove the screws that ONLY hold the top panel
5. Remove the overflow container if you have purchased this accessory with the micro-bath
6. Remove the drain cover and then remove the large drain bolt
7. Remove the overflow bolt
8. Remove the two screws that hold the back panel
9. Lift the top panel up for about an inch and then pull the back panel outwards
10. Now proceed to replace the power entry module