Views:

Overview

To successfully run MET/CAL procedures that use the Fluke 96270A and to take full advantage of its many options used in spectrum analyzer and other RF/µW procedures, the instrument and all required options must be configured properly in both MET/CAL and MET/TEAM.

MET/CAL Configuration

Barcode

Like any MET/CAL reference, begin by entering the basic 96270A information, including the Barcode (Asset Number). The asset number must match an identical, valid entry in the database.

Fluke_96270A_barcode.png

IEEE-488 Address

Enter the IEEE-488 Address, ensuring the value matches the setting in the 96270A GPIB Preferences. The instrument’s GPIB Personality Name should be set to 96270A and its corresponding Address & State fields set to the correct address (18, is the factory default) and Active state.

Figure_1.png

Fluke_96270A_IEEE_address.png

Alias

Enter any required Alias in this field. If it is necessary to execute procedures developed using older model(s) and their respective instrument FSCs, e.g., 9640A, 9640A-LPN, aliases are required and must be added.  In this case, “9640”, “LPN,” or other entries are used for proper accuracy mode identification as shown below. (see the MET/CAL help system for a complete list of applicable model aliases).

Fluke_96270A_Alias.png

Low Level µW Attenuator

If the 96270A has option (96270A/LL) installed, the matching option box must be checked.

Fluke_96270A_Low_Level_Attenuator.png

Auxiliary Device

Next configure the Auxiliary Devices such as the 96040A-50 (50 O) or 96040A-75 (75 O) Leveling Heads, and power splitter models (11667A or 11667B).  Note: Auxiliary Device configuration must include current barcodes (asset numbers) as required for procedure execution. Both models of Leveling Heads and Power Splitters may be configured simultaneously, when applicable.  At least one Leveling Head is required for basic 96270A operation.  When configuring the High Frequency model 96270A/HF, the addition of the 11667B model of power splitter is also required.

Basic Aux Device Configuration

Fluke_96270A_Aux_Device_Basic.png

All Aux Devices Configured

Fluke_96270A_Fully_Configured.png

SENSORS 1 / SENSORS 2

If the instrument includes the 96270A/HF or 96000SNS options, the power sensor(s) must be configured prior to use. SENSORS 1 and SENSORS 2 refer to the 96270A power meter input channel assignment. In the case of the 96270A/HF option, the power sensor (Rohde & Schwarz NRP-Z55.03) is configured by selecting the sensor model from the drop-down list (SENSORS 1, in this example) and adding its barcode entry. Note that two different models of supported sensors may be configured at the same time. If the 96000SNS option (a second, R&S NRP-Z55.03 Power Sensor) is included, (SENSORS 2, in this example) it must be configured using the same method as SENSORS 1, again selecting the correct sensor model from the drop-down list and entering its asset number.

Single Power Sensor Configuration

Fluke_96270A_Sensor_1.png

Dual Power Sensor Configuration

Fluke_96270A_Sensors_1_and_2.png

Assuming the barcodes entered are identical to those in the database and their calibration status is current, the 96270A configuration in MET/CAL is complete.

MET/TEAM Configuration

Adding your 96270A to MET/TEAM requires creating the 96270A Type and Asset records as well as a Scheduled Service to show that the reference is in calibration.

Type

Log into your MET/TEAM webpage:

login.jpg

Click on the Maintenance tab, then select “Type”:

Maintnance_Menu.jpg

On the Find Type Screen, search for an existing Type for the 96270A. If one already exists, select it and skip to the “Asset Record” section.

Find_Type.JPG

Click “Add” on the Find Type screen. This will open the “Add Type” screen.

Add_Ok_Cancel.jpg

Add the Description (RF Reference, for example), Model Number (96270A) and Manufacturer (Fluke). You can enter any other Type details pertinent to your database and click the Save icon.

Add_Type.jpg

Save and “Close” the Add Type screen.

Asset and Scheduled Service

Now we need to add this specific 96270A asset to MET/TEAM. To add an Asset, look at the “Add an Asset” in this article. This article includes instructions on how to create a Scheduled Service, which is required before using this reference with MET/CAL.

Final Steps to Calibrate your DUT

Now that we have added the 96270A asset to MET/TEAM, we can receive the DUT Asset and open a Work Order.  This process will allow you to run the PXE Procedure you associated to the Type in MET/TEAM. To learn how to receive an Asset, review this article.

After creating the Type and Asset for the 96270A, and DUT Work Order you are ready to "Run" your DUT procedure with MET/CAL Runtime.