The purpose of this article is to identify the part number for the cable, see a picture of the cable and provide instructions on replacing the cable.
The part number for the cable is 4468923.
Note that this is a replacement cable for a 750P pressure module. It will not work with a 700P pressure module.
Below is an image of an example 750P pressure module.
The 750P pressure module cable can become damaged due to improperly connecting the Lemo connector to a Fluke pressure calibrator or from wear and tear over the years in a harsh environment.
The good news is that the cable can be replaced in the field without the need for re-calibration.
The 750P cable has a connector on the end connecting to the 750P, no soldering is required.
See the instructions and pictures below for reference.
1) Remove the 4 screws on the back of the 750P pressure module and remove the bottom cover.
2) Replace and route the cable as shown, connecting to the PCA.
3) Replace the back cover and test to verify correct operation.