1-If you are currently using MET/TEMP II version 5.0 you can click on the link below to update to version 5.1. If you are currently using version 4.X you must purchase the upgraded version “9938-V5-UPG (PN: 4699287)", and after purchasing the upgraded version you should receive a new serial number. Do not try to upgrade the old version 4.1 to version 5.1 in the old computer and instead proceed to do a new installation in the new Windows 10 computer using the new serial number. Version 5.1 includes the USB feature.

**If you are installing the full version 5.1 after purchasing the 9938-16-v5 or 9938-25-V5 you can still use the link below to install V5.1 and enter the serial number when prompted. After the software is installed follow steps 7 through 9 to use the USB Hub feature.

https://us.flukecal.com/literature/software-downloads/software/met-temp-ii-software-download

2-In the below dialog click yes.

mceclip0.png

3-Click “Next”.

mceclip1.png

4-Click on “Repair” then “Next”. If you are installing version 5.1 for the first time you won’t get the below dialog.

mceclip2.png

5-Click “Install”

mceclip3.png

6-Click Finish and the software will be now updated to version 5.1

mceclip4.png

7-Launch MET/TEM II, now click on “File” then click on “Defaults…” and finally click on “SmartSwitch”. By default, “None” is already selected, and this will allow you to use a USB Hub.

 

mceclip5.png

Online you can find a variety of USB Hubs such as the one below that would work well. Fluke does not sell the USB Hub.

mceclip6.png

8-An adapter RS-232 to USB is needed to plug the readout and heat sources to the USB Hub.

8.1-Most of our heat sources and readouts will require the below Null Modem RS-232 to USB cable, and our Fluke part number for it is 3525836. It can be purchased by sending an email to orders@flukecal.com for US customers or by contacting your local Fluke authorized distributor for customers outside the US. Click in the link below to search for an authorized distributor.

https://us.flukecal.com/where-to-buy

mceclip7.png

8.2-If you have a reference readout 1594A, 1595A, or 1586A use a USB cable like the one below to connect the unit to the USB Hub and you can purchase it locally. Fluke does not sell this cable.

mceclip8.png

8.3-If you are using the 1523 or 1524 as a reference readout and the 1620A to monitor the temperature or humidity during the calibration we recommend to use a USB to serial stereo cable like the one in the picture below. This cable can be purchased online and below is a link from Amazon. Fluke does not sell this cable.

https://www.amazon.com/RS232-3-5mm-Serial-Adapter-USB-RS232-AJ/dp/B08CQYJSFC/ref=sr_1_5?crid=205SKPDQGHXFW&keywords=usb+to+3.5+mm+serial&qid=1666822568&qu=eyJxc2MiOiIxLjU5IiwicXNhIjoiMC4wMCIsInFzcCI6IjAuMDAifQ%3D%3D&sprefix=usb+to+3.5+mm+serial+%2Caps%2C190&sr=8-5

mceclip9.png

9-Because you are now using a “USB Hub” instead of a smart switch you need to identify which “COM Port” number is the computer using for your reference readout and all the heat sources. Device Manager in your computer is a very useful tool to identify which com ports the computer is using for each of the instruments.

All other settings are identical than when you were using the smart switch.

mceclip10.png