A) Freezing the cell:
- Press the “Set” button and while holding it down, press the “Down” button
- When the display reads “ADV” release the Set and Down buttons and press the Up button repeatedly until the display reads “FREEZC”
- Press the “Set” button once and the freezing process will begin
- When the cell is frozen the display will read “AUTO
B) Realization of the cell:
- The display should read “AUTO”
- Gently insert the Heater into the cell
- Press the “Set” button once
- The 9230 will realize the cell
- After the 9230 beeps eight times and the display reads “MAINT”, the cell is ready to be used
OPTIONAL: The best readings are achieved when there is about 5 centimeters of oil (silicone or mineral) at the bottom of the inner well in the Gallium cell. The silicone oils than can be used are models 5010 or 5012