Applies To:
9144 & 9144-P
Problem Description:
While attempting to heat my 9144 trips my RCD or GFCI.
Resolution/Work Around:
Before initial use, after transport, or after storage in humid or semi-humid environments, the instrument must be energized for a "dry-out" period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 61010-2-010. Use a 2°C/min scan rate and a set point of 100°C.