Resolving SMFT-1000 Detection Issues on Windows

If your PC does not recognize the SMFT-1000 when connected, this guide will help you troubleshoot and resolve the issue.

Requirements

  • Administrative rights on your Windows PC.

  • The SMFT-1000 is powered and set to “Menu” on the rotary dial.

Instructions

Step 1: Install the Fluke IR189USB USB Cable Adapter Driver

Step 2: Verify the Driver Installation

  • After installation, ensure the driver is correctly recognized by your system.

  • Open the Device Manager on your Windows PC.

  • Check for a USB Serial Port entry under Ports (COM & LPT).

  • The driver provider listed in the device's properties should be FTDI.

Step 3: Restart Your PC

  • A system restart can help finalize the installation process.

  • Save any open work and close all applications.

  • Restart your PC to apply the new driver settings.

Step 4: Attempt Firmware Update

  • With the driver installed and the PC restarted, update the firmware using the TruTest software.

Additional Tips

  • Double-check the USB connection if the device is still not recognized.

  • Ensure the SMFT-1000 is correctly powered and set to the appropriate mode.

Support

For further assistance, please visit my.fluke.com to open a support case.