Question:
Although I have put fresh batteries in my Fluke 1503/1507, I get a "low battery" indication when I switch to the Insulation function. Why?
Answer:
This and similar questions have prompted us to test this. Three sets of new Duracell batteries were tested, in several Fluke-1503/1507's with the same result.
Other brands of batteries may cause the same result as well.
The following procedure can be used to turn-off the low battery message ;
- Set the 1503/1507 to V function and press and hold down the BLUE button.
- The tester will indicate first BAT and then the voltage-level of the batteries.
- Turn the unit off and then back on. Observe that the empty battery indicator does not appear when making a insulation measurement.
With this method the high current battery test mode is used.
It puts a load across the batteries that reduces the internal impedance of the batteries to an acceptable level for the battery-check electronics.