Views:

A list of spare Parts that are available to customers is at the end of this article, and is attached as an Excel file.

 

** 1/8in tubing:

Part # 4863095
Description :FLUKE-719P-8003,HOSE, NYLON , 0.125 X 0.064

** Note that the last item, power supply 4312648 (16V) has been replaced by new part number 4769257 (15V). If you are replacing a 16V power supply with a 15V power supply, the charger board needs to be replaced at a Fluke service center. If the charger board is not replaced the 3130 may charge slowly or not fully. Also, see this article if you would like to purchase a 16V power supply on your own, 3130 Power Supply Replacement

*** Also note that the current model battery pack has an additional plug connector on it for a temperature sensor. If your 3130 calibrator does not have the associated jack connector then don't connect the temperature sensor plug and secure it inside the 3130.

 

 

 

Parts can be ordered from orders@flukecal.com or by calling +1.877.355.3225 or contacting your local representative.

The attached Excel file is the same as this image.

** Note that the last item, power supply 4312648 (16V) has been replaced by new part number 4769257 (15V). If you are replacing a 16V power supply with a 15V power supply, the charger board should be replaced, or the 3130 may charge slowly or not fully. Replacement can be done at Fluke Calibration Authorized Service Centers. Contact Fluke Calibration Service at the below link or by calling 877-355-3225